AERI Lights Startup Sequence

Record what the startup sequence is of the 4 lights.

  1. Cycle the power of the instrument off via the power strip
  2. Check the power, scan, zpd & data lights on the interferometer.
    1. You can do this by looking at the reflection of the lights that reflect off the metal plate behind the  (see image below).
  3. Record the sequence the lights cycle through.
  4. At the end of the startup sequence normal operation is Power LED(green top) “ON” and Data LED (yellow, bottom) blinking.  If the laser is dead, the Scan LED(Red, second one from the top) will be “ON”