Interferometer Status Display Code = 3

Symptom

  • Interferometer doesn’t start up normally
  • Scan, ZPD, Data lights are OFF
  • Status display code is 3:  The DSP is searching; the mechanical ZPD (full path bumper to bumper)
  • In open loop mode, FTSW does not initialize:  “Unable to open board”

Possible Causes

  • Vibration from moving or shipping is causing the interferometer mechanism to not scan correctly

Troubleshooting

  • The scan mechanism needs to be physically reset:
  • Power cycle the AERI hardware
  • Run the interferometer in open-loop mode: this operates the scan actuator with increased power
  • Physically move the interferometer around: carefully tilt it up to 90 degrees in each direction