M-AERI Daily Preventative Maintenance Procedure – SAIL
Instrument: M‑AERI
Starting Date: 1 July 2021
Instrument Mentor: Jonathan Gero
Facility: AMF‑M2
PM Frequency: Daily
Hardware checks:
Physical check of instrument:
- Check for any damage
- Check that the 3 TEC fans are operational
- [If hatch open] Clear any debris / water / ice accumulation inside the front-end
- [If hatch open] Check that the sky and side fields-of-view are clear of obstructions (e.g. spider webs, ice)
- [If hatch open] Check that the mirror drum is rotating freely
- [Winter] Check for ice buildup, especially around the hatches
Inspection of hatch state:
- Verify that the hatches are in the correct state (open/closed) appropriate to the precipitation
- If they are not, check on the FTSW display if a rain sensor has been triggered
- outsideRainSensor=0 or insideRainSensor=0 indicates a triggered rain sensor
- Clean the rain sensor with water or alcohol, or troubleshoot as needed
- Note that software outages may also cause the hatch to close (see software troubleshooting)
- If they are not, check on the FTSW display if a rain sensor has been triggered
- Verify that the hatches are either open or closed all the way (apply grease to slots if necessary)
Software checks:
Ingest:
- Check that Ingest is running in the black cygwin window, scrolling, displaying scene numbers (e.g. S42) (see software troubleshooting)
FTSW:
- Acquisition tab: Check the 2 graphs (blue & red) of AERI spectra are displayed and updating (see software troubleshooting)
- Monitoring tab: Check that all status boxes are green (see software troubleshooting)
- Check the interferometer humidity box (replace desiccant if needed)